Skip to main content
. 2017 Aug 1;3(10):475–478. doi: 10.1016/j.hrcr.2017.07.009

Figure 2.

Figure 2

Device interrogation before system repositioning. The primary vector (1× gain) showed less oversensing but an unstable R-wave amplitude owing to postural movements (intermittently <1 mV). Maneuvers resembling the exercise that caused the inappropriate shock lead to intermittent oversensing (asterisks). Small R-wave amplitudes were also observed in alternate and secondary vectors.