. 2017 Oct 17;16:418. doi: 10.1186/s12936-017-2065-8

Table 1.

Shows malaria diagnosis for 115 retrospective samples by the illumigene malaria assay in comparison with standard microscopy, RDT and PCR

Reference method (RDT, microscopy, PCR) Illumigene malaria assay
Results (n) Range parasite density (asexual parasites/µL) Results (n) Performance
Total 108 108
 Positive  74  74 Sensitivity (%) (95% CI) 100% (95.1–100%)
 Negative  34  34 Specificity (%) (95% CI) 100% (89.7–100%)
Clinical samples 96
 Pf  27 (1–372,117)  –*
 Pv  14 (169–18,777)  –*
 Pm  13 (106–9244)  –*
 Po  12 (31–6461)  –*
 Mixed  0
 Negative  30
External quality controls 12
 Pf  1 5  –*
 Pv  1 5  –*
 Pm  2 5  –*
 Po  3 20  –*
 Pk  1 10,000
 Negative  4

The retrospective panel consisted of 103 clinical samples (96 diagnostic samples and 7 samples from patients who already received treatment for malaria infection) and 12 external quality controls. These 7 samples from patients who already received treatment (115 − 7 = 108) were excluded from the calculation of both sensitivity and specificity. Parasite density as quantified by microscopy on blood smear

RDT, rapid diagnostic test; PCR, four-primer real-time PCR; Pf, Plasmodium falciparum; Pm, Plasmodium malariae; Po, Plasmodium ovale; Pv, Plasmodium vivax; Pk, Plasmodium knowlesi; n, number of samples; –*, the illumigene malaria assay does not distinguish between Plasmodium species