Table 2.

Comparisons results showing the difference in pDgNCT coefficients for the cylindrical phantoms relative to the breast phantoms averaged across a range of tube potentials from 35 to 70 kV (1 kV intervals)

Phantom: V1 V3 V5
Beam filtration: 0.2 mm Cu 1.5 mm Al 0.2 mm Cu 1.5 mm Al 0.2 mm Cu 1.5 mm Al
Rcw: Difference (%) −14.2 −16.2 −12.1 −13.1 −12.2 −12.7
RCOM: Difference (%) −0.6 −1.5 0.3 0.1 2.0 2.1