Published in final edited form as: Cornea. 2017 Oct;36(10):1159–1165. doi: 10.1097/ICO.0000000000001294

Table 3.

Inter-rater reliability for the diagnosis of corneal pathology from iTouch and Nidek photo series

Second Grader Diagnosis
First Grader Diagnosis Normal Pathology Kappa (CI)
iTouch
Grader 1 vs. Grader 2
 Normal 94 7 0.54 (0.43, 0.65)
 Pathology 38 59
Grader 1 vs. Grader 3
 Normal 90 11 0.63 (0.53, 0.74)
 Pathology 25 72
Grader 2 vs. Grader 3
 Normal 110 22 0.71 (0.61, 0.81)
 Pathology 5 61
Nidek
Grader 1 vs. Grader 2
 Normal 104 8 0.75 (0.66, 0.84)
 Pathology 16 70
Grader 1 vs. Grader 3
 Normal 96 16 0.76 (0.66, 0.85)
 Pathology 8 78
Grader 2 vs. Grader 3
 Normal 100 20 0.76 (0.66, 0.85)
 Pathology 4 74

CI = 95% Confidence Interval