Table 2.

Comparison of the three systems

AIRVO 2
Constant flow
Steady air
Constant flow
Vivo 50
Constant pressure
p Values
Difference in tidal volume (mL) 221.40±263.92 257.30±315.06 244.24±268.63 p>0.99
Difference in PtcCO2 (mm Hg) −0.50±2.37 −0.30±1.06 −0.60±2.32 p>0.05
Difference in RR (bpm) −3.26±2.14 −3.17±4.93 −4.18±2.68 p>079
Difference in EIOP (cmH2O) −3.18±4.70 −2.04±1.74 −1.77±2.56 p>0.85
Difference in EEOP (cmH2O) −0.51±2.80 0.71±2.34 1.02±3.99 p>0.92

p Values in comparison between NHF-OTS 30 L/min and baseline.

difference EEOP, difference in end-expiration oesophageal pressure; difference EIOP, difference in end-inspiration oesophageal pressure; PtcCO2, transcutaneous carbon dioxide; RR, respiratory rate.