Table 1.
The specifications of the analyzer parameters used for MultiFLEXX. All analyzers are composed of 3 HOPG crystals of 20 × 20 mm2 cross section.
| Channel | 2.5 meV | 3 meV | 3.5 meV | 4 meV | 4.5 meV |
|---|---|---|---|---|---|
| [mm] | 1050 | 1220 | 1387 | 1552 | 1732 |
| [1 /m] | 0.76 | 0.6 | 0.5 | 0.41 | 0.37 |
| [ eV] | 53 | 78 | 103 | 130 | 157 |
| [ eV] | 89 | 106 | 123 | 140 | 164 |
| take-off angle [deg.] | 116.98 | 102.20 | 92.20 | 84.74 | 78.90 |
| [] | 1.1 | 1.2 | 1.3 | 1.39 | 1.47 |
| [a.u.] | 1 | 0.68 | 0.48 | 0.36 | 0.27 |
| [a.u.] | 1 | 1.70 | 2.57 | 3.59 | 4.71 |
| Normalization factor | 1 | 1.16 | 1.23 | 1.30 | 1.27 |
The elastic linewidths were simulated using McStas, and the tabulated measured elastic linewidths are averages over all analyzers with the same . The curvature of the analyzer segments is given by ρ. The covered solid angle has been calculated from the sample-analyzer distance and the take-off angle . The resolution volume is calculated directly from the take-off angle and the final wavevector . Both variables are normalized to the values for the meV analyzer.