Skip to main content
. 2017 Jun 16;8(44):76076–76084. doi: 10.18632/oncotarget.18530

Table 3. Changes in gamma passing rates of the failed cases by simulating setup errors.

Gamma passing rate with 2%/1 mm (%)
Couch rotation –0.2° –0.1° 0.1° 0.2°
T-spine 84.6 84.6 84.6 84.6
Lung 89.4 89.5 87.2 87.2
L-spine 87.3 87.9 87.8 87.5
C-spine 88.6 88.6 88.4 88.9
SI shift –0.2 cm –0.1 cm 0.1 cm 0.2 cm
T-spine 83.0 84.6 84.6 84.6
Lung 82.2 89.0 89.0 85.5
L-spine 88.4 87.5 87.5 86.0
C-spine 88.0 88.7 88.7 86.2
LR shift –0.2 cm –0.1 cm 0.1 cm 0.2 cm
T-spine 82.7 84.6 84.6 84.1
Lung 97.5 100 87.6 85.5
L-spine 90.4 91.6 87.5 85.5
C-spine 85.5 88.7 88.7 86.0
AP shift –0.2 cm –0.1 cm 0.1 cm 0.2 cm
T-spine 73.1 79.2 57.6 24.3
Lung 64.0 73.3 55.6 54.5
L-spine 82.6 86.2 85.6 81.4
C-spine 81.7 84.7 94.2 93.2

Abbreviations: SI, superior to inferior direction; LR, left to right direction; AP, anterior to posterior direction

Note: The gamma passing rates higher than 90% are shown in italic.