Top: Acquisition parameters for the 3T Siemens (Si) Skyra and Prisma scanners. Bottom: Test-retest reliabilities for selected ROIs, processed by Freesurfer. The ROIs are gray matter volume (GMV), subcortical gray matter volume (scGMV), cortex volume (cVol), cortical white matter volume (cWMV), and the volumes of the lateral ventricle (LV), amygdala (Amyg), thalamus (Thal), hippocampus (Hipp), caudate (Caud), and finally the estimated total intracranial volume (eTIV). Test-retest reliability is computed as within-site ICC(1,1)

10 11 12 13 14 15
TR (ms) 2300 2300 2300 2300 2300 2000
TE (ms) 2.96 2.98 2.98 2.96 2.96 3.22
Strength (T) 3 3 3 3 3 3
TI (ms) 900 900 900 900 900 900
Flip Angle (°) 9 9 9 9 9 8
Make Si Si Si Si Si Si
Voxel Size (mm) 1×1×1 1×1×1.1 1×1×1 1×1×1 1×1×1 1×1×1
Distortion Correction Y N Y Y Y N
Parallel Imaging 2 - 2 2 2 2
FOV (mm) 256×256×176 240×256×176 240×256×176 240×276×156 256×256×176 256×208×160
Read Out Direction HF RL HF HF HF RL
Head coil # channels 20 32 20 20 20 32
Model Prisma Prisma fit Skyra Skyra Skyra Skyra
OS D13D VD13D VD13 VD13 VD13C VD13
Acq. Time (min) 05:09 07:46 05:12 05:12 05:09 04:56
orientation sag sag sag sag sag ax
# scans 22/22 24/24 25/25 23/24 23/24 22/22

Amyg (L) .83 .89 .80 .85 .98 .89
Amyg (R) .94 .92 .93 .85 .93 .84
Caud (L) .99 .99 .98 .99 .98 .98
Caud (R) .99 .96 .95 .95 .98 .97
GMV .99 .98 .99 1 .99 .97
Hipp (L) .94 .98 .99 .95 .97 .98
Hipp (R) .91 .94 .97 .98 .95 .96
Thal (L) .92 .87 .87 .76 .91 .89
Thal (R) .74 .93 .80 .91 .93 .89
cVol .99 .98 .98 1 .99 .96
cWMV 1 1 1 1 1 .97
eTIV 1 1 1 1 1 .97
scGMV .98 .99 .98 .98 .99 .99