Table 2.
Common problems and solutions to fluid flow related errors. Those problems or solutions marked with an asterisk* are specifically related to the example system. All other problems and solutions are general troubleshooting strategies.
| Problem | Solution |
|---|---|
| Air bubbles | Manually purge the entire flow path with buffer |
| Seal/fluid junction leak | Adjust connection, reinforce with Teflon tape, or replace connection |
| Clog in cartridge top* or microfluidic component | Remove component, sonicate if possible, manually purge with multiple solvents, and reassemble |
| Clog in tubing | Replace tubing (do not force through pumps) or separate from set up and thoroughly purge with multiple solvents |
| Broken seal between cartridge top and chip* | Remove cartridge top, dry, carefully reassemble* |
| Reduced or loss of signal after initial troubleshooting | Run a reliable test assay (Instrument calibration) |