TABLE 1. Accuracy of fluorescence microscopy and glutaraldehyde (GA) test using a culture reference standard.

All specimens N = 272
Methods Culture Sensitivity (%) 95% CI Specificity (%) 95% CI PPV (%) 95% CI NPV (%) 95% CI
Positive Negative
Fluorescence microscopy Positive 56 0 57.1 46.8-67 100 97.3-100 100 92-100 80.6 74.5-85.5
Negative 42 174
GA test Positive 93 0 94.2 87.9-98.1 80.5 73.6-85.9 73.2 64.5-80.5 96.8 91.7-98.7
Negative 5 174
Respiratory specimens, N = 176
Fluorescence microscopy Positive 42 0 65.6 52.6-76.8 100 95.9-100 100 89.6-100 83.6 76-89.2
Negative 22 112
GA test Positive 62 0 96.9 88.2-99.5 82.1 73.5-88.5 75.6 64.7-84.1 97.9 91.8-99.6
Negative 2 112
Non respiratory specimens, N = 96
Fluorescence microscopy Positive 14 0 41.2 25.1-59.2 100 92.7-100 100 73.2-100 75.6 64.7-84.1
Negative 20 62
GA test Positive 31 0 91.2 75.2-97.7 77.4 64.7-86.7 68.9 53.2-81.4 94.1 82.8-98.5
Negative 3 62

CI: confidence interval; NPV: negative predictive value; PPV: positive predictive value.