Skip to main content
. Author manuscript; available in PMC: 2018 Oct 1.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2017 Jul 11;64(10):1439–1449. doi: 10.1109/TUFFC.2017.2725839

TABLE IV.

Measured safety parameters with three Δϕ’s (0°, 140°, and 180°)

Δϕ MI ISPTA (mW/cm2) ISPPA (W/cm2)
1.73 70.49 117.48
140° 0.77 11.40 19.00
180° 0.72 9.45 15.76

FDA limit 1.9 720 190