Table 1.
Pairwise comparative analyses between each NCT and GAT
NCT device | Mean (±SD) (mmHg) | Mean difference (±SD) with G-IOP | 95% LoA | p value |
---|---|---|---|---|
C-IOP | 17.23 ± 2.94 | 1.38 ± 1.75 | −2.05 to 4.82 | <0.001 |
N-IOP | 14.87 ± 3.25 | −0.98 ± 2.00 | −4.91 to 2.95 | 0.001 |
T-IOP | 16.33 ± 3.01 | 0.48 ± 2.12 | −3.67 to 4.63 | 0.108 |
C-IOP IOP value measured by Canon NCT, N-IOP IOP value measured by Nidek NCT, T-IOP IOP value measured by Topcon NCT, G-IOP IOP value measured by GAT, LoA limits of agreement
P value by paired t-test