Table 3. Summary of results from manufacturer provided daily quality control material (ImmunotrolTM Normal and Low) for 12 consecutive days, where target value and range refers to package insert values.
Normal ImmunotrolTM | Low ImmunotrolTM | |||
---|---|---|---|---|
CD4# | CD4% | CD4# | CD4% | |
Number of values | 12 | 12 | 12 | 12 |
Target value (package insert) | 569±91 | 41.1±3.7 | 136±33 | 15.3±3.3 |
Target Range | 478–660 | 37.4–44.8 | 103–169 | 12.0–18.6 |
Tested Range | 550–620 | 39.9–41.5 | 122–157 | 15.0–17.1 |
Median ±Standard Deviation | 573.7±24.63 | 41.47±0.99 | 139.4±10.62 | 15.79±0.60 |
95% Confidence Interval of Mean | 558–589.3 | 40.84–42.11 | 132.7–146.2 | 15.41–16.17 |
%Coefficient of variation (precision) | 4.29 | 2.41 | 7.62 | 3.82 |