Acceptance criteria for standards, controls, and test samples

Standards & Controls Acceptable Range (GC/5 μl)Low to High Acceptable %CV
PEC 1.58 × 103 1.58 × 104 ≤ 40%
50 HA Standard 1.58 × 101 1.58 × 102 ≤ 40%
500 HA Standard 1.58 × 102 1.58 × 103 ≤ 30%
5000 HA Standard 1.58 × 103 1.58 × 104 ≤ 30%
50,000 HA Standard 1.58 × 104 1.58 × 105 ≤ 30%
500,000 HA Standard 1.58 × 105 1.58 × 106 ≤ 30%
5,000,000 HA Standard 1.58 × 106 1.58 × 107 ≤ 30%
PTS NA NA ≤ 30%
Efficiency 1.8 2.2 < 10%

PEC positive extraction control

PTS positive test sample