Skip to main content
. Author manuscript; available in PMC: 2017 Nov 6.
Published in final edited form as: Anal Chem. 2016 Jun 27;88(14):7302–7311. doi: 10.1021/acs.analchem.6b01655

Table 3.

Measurements of Spatial Resolution Using Line Gratings across Various Instrument Settingsa

laser focus setting laser power (%) smallest resolved separation (μm)
single 19.0 <30
21.0 <30
23.0 <30
Mod 5 39.4 50
43.4 60
47.4 60
Mod 5 defocused 62.8 <30
66.0 35
70.0 70
a

Data were collected using a Bruker Rapiflex.