Table 6. Device Related Complications in the Study Population.

Patient Indication Age at Implantation (years) Implant to Death (years) Age at Death (years) Time to Appropriate Therapy (days) Time to Inappropriate Therapy (days) +/- 1 day of implant < 30 Post Implant Complication > 30 Day Post Implant Complication Complication
209 Primary 73.2 0.8 74 Yes Hematoma
133 Secondary 72.7 2.3 75 Yes Infection
258 Primary 83.9 3.1 87 Yes Lead fracture leading to multiple shocks
241 Primary 70.9 5.1 76 791 Yes Recall
254 Primary 72.3 5.7 78 Yes Vegetation on lead
131 Primary 77.3 6.7 84 63 Yes Hematoma
194 Primary 79.1 6.9 86 497 1654 Yes Recall
267 Primary 72.2 7.8 80 53 Yes LV lead dislodgement
182 Primary 74.6 8.4 83 Yes Device recall
193 Secondary 75.6 8.4 84 1440 Yes Lead Fracture
66 Secondary 78.3 8.7 87 Yes Recall-component failure
273 Secondary 70.6 9.4 80 191 2870 Yes Lead fracture
260 Secondary 76.8 14.2 91 4473 Yes Lead revision
41 Primary 77.4 Yes Hematoma
54 Secondary 76.6 Yes Battery recall
55 Primary 86.4 669 Yes Battery recall
65 Primary 73.1 Yes ICD pocket infection
101 Primary 75.7 Yes LV lead noise
102 Primary 81.1 Yes Infection
105 Primary 77.6 Yes CS Dissection
106 Secondary 72.4 2424 Yes Lead fracture
122 Secondary 73.2 Yes RV perforation
129 Primary 72.2 Yes Hematoma
141 Primary 75.9 306 Yes Battery Recall
160 Primary 80.3 Yes Hematoma