Table 1. . A summary of the electrical biosensors based on ZnO for detection of cardiac biomarkers.
Cardiac biomarker | ZnO nanostructures type | Detection format | Detection range | Lower detection limit | Ref. |
---|---|---|---|---|---|
cTnT | Nanorods | Immunosensor | 1 fg/ml to 100 ng/ml | 100 fg/ml | [72] |
cTnT | Nanorods | Immunosensor | 1 fg/ml to 100 ng/ml | 1 pg/ml | [74] |
cTnI | Nanoparticles | Electrical | 1 ng/ml to 10 μg/ml | 2.19 ng/ml | [71] |
cTnI | Nanoparticles | FET | 1 ng/ml to 10 μg/ml | 3.24 pg/ml | [72] |
cTnT & cTnI | Nanorods | Immunosensor | 100 fg/ml to 100 ng/ml | 1 pg/ml | [73] |
FET: xxx.