Table 1.

Assay accuracy and variability results for acyclovir, ganciclovir, and H2G from five validation runs

Quality control (ng/mL) Acyclovir (%) Ganciclovir (%) H2G (%)
Accuracy
50 106.4 99.0 102.0
100 100.5 92.1 94.1
1000 91.7 89.5 92.5
10,000 92.2 92.5 93.6
Within assay variability
50 4.8 1.6 4.8
100 2.6 1.9 2.6
1000 1.7 1.6 2.0
10,000 0.8 1.0 0.7
Between assay variability
50 7.9 4.5 3.7
100 4.7 2.8 1.9
1000 2.3 2.7 2.7
10,000 1.1 1.6 1.0

CV, coefficient of variation.