Published in final edited form as: J Magn Reson. 2017 Feb 14;277:45–51. doi: 10.1016/j.jmr.2017.02.009

Table 1.

Resonator and sample cell characteristics calculated and measured. EPR signals are at constant microwave field and normalized to a 1 mm capillary in a cylindrical TE011, while Λave can be directly compared.

Geometry freq. [GHz] Q-value Norm. Signal Λave [G/W1/2] Vol. [μL]
Meas. Calc. Meas. Calc. Meas. Calc. Meas. Calc.

Two-Loop–One-Gap

TPX (0.6 mm I.D.) 9.23 9.34 204 268 0.21 0.32 5.61 4.37 2.2
DoubleDee Par. 9.31 9.41 419 408 0.31 0.43 6.83 5.32 3.2
DoubleDee Perp. 9.29 9.38 387 352 0.25 0.39 5.82 4.96 3.2

Five-Loop–Four-Gap

Capillary (1 mm I.D.) 9.56 9.75 620 802 0.35 0.35 1.67 1.78 7
AquaStar 9.44 9.63 673 640 0.83 0.81 1.52 1.61 25

Cylindrical TE011

Capillary (1 mm I.D.) 9.86 9.58 3651 6640 1 1 1.12 1.45 30
AquaX 9.86 9.53 2466 4040 3.24 3.17 0.52 0.47 143
AquaStar 9.86 9.58 2530 5154 3.02 2.97 0.68 0.78 100
AquaSun 9.43 1848 8.18 0.52 396