Table 5.

Performance of the EC sensor in the lab (n = 3). In parenthesis are the simultaneous reference measurements.

Platform Averaging Time (min) Pollutant Mean ± STD at Zero-Air (ppb) Mean ± STD at 100 ppb span * (ppb) R2 Gain Intercept (ppb) Cross-Sensitivity
AQMesh 15 CO 16.3 ± 6 (1.9 ± 0.7) 1292 ± 21.5 (1385 ± 16.2) 0.99 0.86 0.07 NO2
NO n/a (0.4 ± 0.4) 88.5 ± 1.5 (94.1 ± 0.9) 0.99 0.97 −1.13
NO2 n/a (0.7 ± 0.3) 126.4 ± 3.5 (103.9 ± 0.7) 0.99 1.22 −1.02
O3 n/a (0.8 ± 0.2) 123.4 ± 2.3 (108.5 ± 1.5) 0.99 1.16 −1.27
LEO 1 NO2 15.3 ± 10.8 (0.4 ± 0.3) 49.0 ± 8.7 (94.3 ± 0.6) 0.99 0.86 23.9 NO2
NO 24.7 ± 3.1 (0.3 ± 0.2) 117.9 ± 3.3 (107.7 ± 0.4) 0.99 0.71 −21.5
O3 6.8 ± 4.1 (0.5 ± 0.5) 57.5 ± 3.4 (86.1 ± 0.6) 0.96 0.70 −7.7

* Except for CO, where 1300 ppb was used as the span value.