Skip to main content
. Author manuscript; available in PMC: 2018 Jun 1.
Published in final edited form as: Clin Biochem. 2017 Apr 12;50(9):513–520. doi: 10.1016/j.clinbiochem.2017.04.006

Table 1.

Experimental design, specimens used, laboratory analysis conducted, findings, and conclusions from the measurement of hemoglobin using the HemoCue® Hb-201+ and/or Hb-301 analyzers1

Experiment Design Specimens Laboratory analysis Finding Conclusion
Comparability of HemoCue® models, including supplies and operation
 1. HemoCue® model comparability Test multiple analyzers of 2 HemoCue® models Venous EDTA blood (n = 25) from TNBS donors Hb-201+ (2 analyzers, 1 reading/analyzer)
Hb-301 (12 analyzers, 1 reading/analyzer)
Hb-301 measured 2.6% higher than Hb-201+ Results from Hb-201+ and Hb-301 are not comparable
 2. Cuvette lot-to-lot variability Test multiple cuvette lots for 2 HemoCue® models Venous EDTA blood (n = 25) from TNBS donors Hb-201+ (2 lots, 1 reading/lot)
Hb-301 (3 lots, 1 reading/lot)
Low variability for both HemoCue® models (CV = 1%) Cuvette lot variability is acceptable
 3. Delayed Hb reading Read Hb at 0 (reference), 1, 2, 3, 4, 5, and 6 min after filling the cuvette Venous EDTA blood (n = 10) from TNBS donors Hb-201+ (1 reading)
Hb-301 (1 reading)
Hb-201+ showed comparable results up to 6 min
Hb-301 showed increasing results (1.3% per min)
Reading on Hb-301 must be carried out within 1 min of filling the cuvette
Robustness of HemoCue® analyzers and supplies
 4. Exposure of analyzers or supplies to elevated temperature a. Keep analyzers at 37°C (incubator) for up to 3 wk and test weekly (reference: 0 wk) Controls (level 1, 2, and 3) for each model Hb-201+ (2 analyzers, 2 readings/analyzer)
Hb-301 (2 analyzers, 2 readings/analyzer)
Keeping analyzers (both models) at 37°C for up to 3 wk did not affect results by more than ±1% Analyzers for both HemoCue® models can withstand elevated temperature
b. Keep cuvettes at 37°C (incubator) for up to 3 wk and test weekly (reference: 0 wk) Venous EDTA blood (n = 24) from TNBS donors Hb-201+ (2 readings)
Hb-301 (2 readings)
Keeping cuvettes (both models) at 37°C for up to 3 wk did not affect results by more than ±1% Cuvettes for both HemoCue® models can withstand elevated temperature
c. Keep Hb-301 controls at 32°C (incubator) for up to 3 wk and test weekly (reference: 0 wk) Controls (level 1, 2, and 3) Hb-301 (2 analyzers, 1 reading/analyzer) Keeping controls (Hb-301) at 32°C for up to 3 wk did not affect results by more than ±1% Controls for the Hb-301 can withstand elevated temperature
 5. Exposure of cuvettes to elevated temperature and humidity a. Keep cuvettes (Hb-301) at 37°C/72% humidity (incubator) in either closed or open boxes for up to 3 wk and test at 0 (reference), 1, 8, 15, and 21 d Controls (level 1, 2, and 3) Hb-301 (2 readings) Keeping cuvettes for Hb-301 (closed or open box) at 37°C and 72% humidity for up to 3 wk did not affect results Cuvettes for Hb-301 can withstand elevated temperature and humidity for several weeks
b. Keep cuvettes (Hb-201+) at 37°C/72% humidity (incubator) in closed boxes for up to 3 wk and test at 0 (reference), 1, 8, 15, and 21 d Controls (level 1, 2, and 3) Hb-201+ (2 readings) Keeping cuvettes for Hb-201+ (closed box) at 37°C and 72% humidity for up to 3 wk did not affect results Cuvettes for Hb-201+ must be stored in the tightly closed box
c. Keep cuvettes (Hb-201+) at 37°C/72% humidity in open boxes for up to 1 h and test at 0 (reference), 10, 20, 25, 30, 40, and 60 min Controls (level 1, 2, and 3) Hb-201+ (2 readings) Keeping cuvettes for Hb-201+ (open box) at 37°C and 72% humidity for up to 1 h produced significantly lower results after 10 min compared to 0 min Cuvettes for Hb-201+, when exposed to humidity, they degrade within 10 min
Venous vs. capillary blood samples
 6. Comparability of venous and capillary blood Measure paired venous and capillary blood samples Venous and capillary EDTA blood (n = 33) from CDC donors Hb-201+ (3 readings) Hb-201+ produced slightly higher results for capillary blood samples, but the difference was not significant In apparently healthy individuals, venous and capillary blood samples produce comparable results when sampled by an experienced technician
 7. Comparability of individual drops vs. pooled capillary blood Measure 2nd, 3rd, and 4th drop capillary blood and compare to pooled capillary blood from Microtainer Capillary EDTA blood (n = 32) from CDC donors Hb-201+ (1 reading) Hb-201 produced comparable results for second or third drop relative to pooled blood; however, fourth drop produced significantly lower results by 3.2% Individual drops of blood (second and third) and pooled blood produce comparable results when sampled by an experienced technician
Sample storage and freeze-thawing
 8. Storing blood at different temperatures for up to 4 d Store venous blood at 10°C, room temperature, and 30°C (incubator) for up to 4 d and test daily (reference: 0 d) Venous EDTA blood (n = 24) from TNBS donors Hb-201+ (2 analyzers)
Hb-301 (2 analyzers)
Hb-201+ produced comparable results at each temperature and day
Hb-301 produced increasing results with time and temperature; storage of blood for more than 1 d at higher than ambient temperature resulted in >1% increase
When using Hb-201+, blood can be stored for a few days prior to measurement at a wide range of temperatures (10–30°C)
When using Hb-301, blood can to be stored at ambient or refrigerated temperature for maximum 1 d prior to measurement
 9. Freeze-thawing a. Freeze capillary blood at −70°C for 4 d and compare to result obtained with fresh blood Capillary EDTA blood (n = 32) from CDC donors Hb-201+ (2 readings) Hb-201+ produced slightly lower results (1.7%) in thawed compared to fresh blood samples Hb should be measured in fresh samples; Hb measured in thawed samples is no longer accurate
b. Freeze venous blood at −70°C, conduct 1 freeze-thawing cycle on day 1 and a second freeze-thawing cycle on day 2 and compare to result obtained with fresh blood Venous EDTA blood (n = 25) from TNBS donors Hb-201+ (2 analyzers)
Hb-301 (2 analyzers)
Hb-201+ produced slightly lower results after 1 (1.8%) and 2 (1.6%) freeze-thawing cycles compared to fresh blood samples
Hb-301 produced slightly higher results after 1 (1.6%) and 2 (2.1%) freeze-thawing cycles compared to fresh blood samples
Hb should be measured in fresh samples; Hb measured in thawed samples is no longer accurate
1

Hb, hemoglobin; TNBS, Tennessee Blood Services