Published in final edited form as: Clin Biochem. 2017 Apr 12;50(9):513–520. doi: 10.1016/j.clinbiochem.2017.04.006

Table 3.

Effect of suboptimal cuvette storage conditions for up to 21 d on measured hemoglobin concentrations (g/L): mean (SD) from 2 readings each

HemoCue® model HemoCue® controls Cuvette condition Time, d
0 1 8 15 21
Hb-301 Level 1 Referencea 71.5 (0.7) 72.0 (1.4) 72.5 (0.7) 72.5 (0.7) 72.5 (0.7)
37°C/humid, closedb 71.5 (0.7) 72.5 (0.7) 71.0 (0) 71.0 (0) 71.5 (0.7)
37°C/humid, openc 71.5 (0.7) 72.0 (0) 72.5 (0.7) 72.5 (0.7) 73.0 (0)
Level 2 Referencea 130 (2.1) 131 (0) 132 (0.7) 130 (0.7) 131 (0)
37°C/humid, closedb 130 (2.1) 131 (0) 130 (0.7) 130 (0) 130 (0.7)
37°C/humid, openc 130 (2.1) 131 (0) 132 (0.7) 132 (0.7) 132 (0)
Level 3 Referencea 171 (0) 172 (0) 175 (0) 174 (0) 175 (0)
37°C/humid, closedb 171 (0) 174 (2.1) 172 (2.1) 171 (1.4) 170 (0.7)
37°C/humid, openc 171 (0) 174 (0.7) 173 (0) 172 (0) 174 (0.7)
Hb-201+ Level 1 Referencea 79.5 (2.1) 78.0 (0) 79.5 (0.7) 79.0 (0) 80.0 (1.4)
37°C/humid, closedb 79.5 (2.1) 80.0 (1.4) 80.0 (1.4) 80.0 (0) 80.2 (1.6)
Level 2 Referencea 116 (0.7) 116 (0) 116 (0.7) 116 (1.4) 120 (0)
37°C/humid, closedb 116 (0.7) 117 (0) 118 (0) 118 (0.7) 120 (3.5)
Level 3 Referencea 158 (0.7) 157 (0.7) 158 (0) 158 (0.7) 160 (2.1)
37°C/humid, closedb 158 (0.7) 157 (0) 159 (0.7) 159 (1.4) 159 (0)
a

Cuvettes were stored in their original box at room temperature in a climate controlled laboratory

b

Cuvettes were stored in their original box in an incubator set to 37°C and 72% humidity

c

Cuvettes were stored in an open container in an incubator set to 37°C and 72% humidity