Table 4.

Effect of suboptimal cuvette storage condition for up to 1 h on measured hemoglobin concentrations (g/L) using the Hb-201+ HemoCue® model: mean (SD) from 2 readings each

HemoCue® controls Cuvette condition Time, min
0 10 20 25 30 40 60
Level 1 37°C/humid, opena 81 (0) 79.5 (0.7) 74.5 (3.5) 79.5 (2.1) 77.5 (0.7) 72.5 (3.5) 74.5 (3.5)
Level 2 37°C/humid, opena 119 (0) 118 (3.5) 102 (4.2) 112 (10.6) 112 (19.1) 110 (21.9) 112 (10.6)
Level 3 37°C/humid, opena 161 (0) 162 (6.4) 145 (2.8) 141 (12.7) 118 (2.1) 113 (1.4) 122 (3.5)
a

Cuvettes were stored in an open container in an incubator set to 37°C and 72% humidity