ACRdelay = Time position of AC with respect to the R peak (in ms)
|
AFPRdelay = Average FPRdelay estimated in the 5 beats preceding or following the current beat |
AO = Opening of the aortic valve (fiducial point) |
MC = Closure of the mitral valve (fiducial point) |
MO = Opening of the mitral valve (fiducial point) |
CFTP = Coarse FP Time Position: FP timing with a 5
ms precision (in ms)
|
D = D1 + D2 |
D1 = Distance (in amplitude) between IRP and the adjacent left minimum (in mg), see Fig. 5 panel c |
D2 = Distance (in amplitude) between IRP and the adjacent right minimum (in mg), see Fig. 5 panel c |
FP = Fiducial Point (either AC, AO, MC or MO) |
FPRdelay = Time position of the generic FP with respect to the R peak (in ms)
|
HRFTP = HiRes FP Time Position: FP timing with 1
ms precision (in ms)
|
ICP = Isovolumic Contraction Point (anchor point used for detection of MC and AO) |
|ICPd| = Absolute value of the ICP distance from the 0 g baseline (in mg) |
ICP
Rdelay = Time position of ICP with respect to the R peak (in ms)
|
ICP
Rdelay
ref = ICP
Rdelay
observed in the closest valid beat (the reference). |
IRP = Isovolumic Relaxation Point (anchor point in S2 used for detection of AC and MO) |
IRP
Rdelay = Time position of IRP with respect to the R peak (in ms)
|
IRP
Rdelay
ref = IRP
Rdelay
observed in the closest valid beat among the last 20 beats (the reference). |
RRI = RR Interval (in ms)
|
S1 = Peak in the SCG envelope corresponding to the first heart sound |
S1si = 50 ms SCG segment starting after 25 ms from the R peak where ICP, MC and AO are searched for |
S2 = Peak in the SCG envelope corresponding to the second heart sound |
S2si = 60 ms SCG segment centered on Te where IRP, AC and MO are searched for |
Te = End of T wave |
Te
Rdelay = Time position of Te with respect to the R peak (in ms)
|