Table 1.
Assignment of ion collectors for isotope analyses of heavy noble gases. All Kr and Xe isotopes are measured simultaneously in only four steps of magnet field and zoom lens, with Kr being analyzed twice using adjacent detectors for internal calibration of counting efficiencies. No changing of magnetic field and zoom lens are needed for Ar analyses.
| Magnet, Zoom lens settings | Ion collectors (EM – electron multipliers, FC – Faraday Cup)
|
||||||||
|---|---|---|---|---|---|---|---|---|---|
| FC | EM1 | EM2 | EM3 | EM4 | EM5 | EM6 | EM7 | EM8 | |
| B1, Z1 | 136Xe | 134Xe | 132Xe | 130Xe | 128Xe | 126Xe | 124Xe | ||
| B2, Z2 | 131Xe | 129Xe | |||||||
| B3, Z3 | 86Kr | 84Kr | 83Kr | 82Kr | 80Kr | ||||
| B4, Z4 | 86Kr | 84Kr | 83Kr | 82Kr | 80Kr | 78Kr | |||
| B5, Z5 | 40Ar | 38Ar | 37Cℓ | 36Ar | |||||