. 2017 Mar 20;18(3):16–27. doi: 10.1002/acm2.12061

Table 3.

Variations in the longitudinally averaged dose around the circumference of the MCVC applicator calculated using TG‐43, sACE, sACE‐20 mm, hACE, and measured using film. Percentages are relative to the mean dose and are given in parentheses. Expanded (k = 2) uncertainties are given

Channel/s Material in grooves Calculation/measurement method Maximum Dose variation [cGy] (%) Mean Dose (cGy) Dose calc. time (min) No. of transport directions (primary/residual)
Central Air sACE 20 ± 66 (4 ± 14) 467 ± 46 3 320/180
Central Air sACE‐20 mm 38 ± 68 (8 ± 14) 491 ± 50 4 320/180
Central Air hACE 37 ± 70 (8 ± 14) 488 ± 48 35 720/240
Central N/A TG‐43 2 ± 48 (0 ± 10) 499 ± 35 N/A N/A
Central Air Film 77 ± 40 (16 ± 8) 477 ± 28 N/A N/A
Central Water sACE 18 ± 68 (4 ± 14) 478 ± 48 3 320/180
Central Water hACE 15 ± 70 (3 ± 14) 484 ± 48 35 720/240
Central Water Film 49 ± 18 (11 ± 4) 464 ± 16 N/A N/A
Peripheral Air sACE 52 ± 68 (11 ± 14) 485 ± 48 28 240/128
Peripheral Air hACE 52 ± 68 (11 ± 14) 482 ± 48 286 500/200
Peripheral Air TG‐43 36 ± 48 (7 ± 10) 500 ± 35 N/A N/A
Peripheral Air Film 61 ± 38 (13 ± 8) 476 ± 30 N/A N/A
Peripheral Water sACE 49 ± 68 (10 ± 14) 487 ± 48 28 240/128
Peripheral Water hACE 50 ± 68 (10 ± 14) 483 ± 48 286 500/200
Peripheral Water Film 57 ± 26 (12 ± 6) 477 ± 14 N/A N/A