Normal tolerances and statistical distributions for various sources of errors of the Varian LINAC machines installed in our clinic

Errors Gantry MU Jaw Collimator MLC
Tolerance 1 MU 1 mm 2 mm
Distribution Uniformly Uniformly Uniformly Uniformly Gaussian