Table 4.

Mobius calculated patient dose vs. AcurosXB for first 36 intensity‐modulated casesa post‐clinical implementation

Target mean dose percent difference (%) Target D95% percent difference (%) 3D global gamma pass rate (%) Target gamma pass rate (%) Dose calculation point percent difference (%)
Average 0.0b −2.0b 99.2c 96.5c 1.8
Max 2.4 4.6 100.0 100.0 4.7
Min −4.8 −10.0 96.7 69.0 −1.6
Standard deviation 1.6 2.4 0.8 5.6 1.3
a

Analyzed cases include first 36 IMRT and VMAT plans treated post‐implementation of the Mobius system in our clinic.

b

A positive percent difference indicates that the Mobius calculated dose is higher than AcurosXB.

c

3D global gamma and target gamma were calculated with a 5%, 2 mm criteria.