Table 3.

Gamma score (1%/1 mm; 2%/1 mm) for dose distribution for static beam delivery of segmented and non‐segmented beam for 20 MU and 200 MU measured at 6 MV with open field of 20 cm by 20 cm

20MU 10MU × 2 4MU × 5 2MU × 10
1% 99.88 55.59 24.87
2% 100 88.42 52.41
200 MU 100 MUx2 40 MUx5 20 MUx10
1% 100 100 93.29