Skip to main content
. 2017 Jan 3;18(1):90–95. doi: 10.1002/acm2.12020

Table 4.

Gamma score (1%/1mm; 2%/1mm) for gated beam delivery for patient A and B using selected beam‐on/off time delivery. Gated beam delivery for FB mode with the beam‐on/off time of (1:3), (1:5), (2:2), and (3:1) and for BH mode with the beam‐on/off time of (12:6)

Duty cycle (%) 50 25 37.5 20 17 75 50 50 38 83 63 50 66.6
Time (s) (1:1) (1:3) (1.5:2.5) (1:4) (1:5) (3:1) (2:2) (3:3) (3:5) (5:1) (5:3) (5:5) (12:6)
1% 99.82 95.72 97.43 96.57 69.42 99.91 99.56 96.62 96.43 99.93 98.27 97.69 100
2% 100 99.69 99.92 100 99.47 100 100 99.96 99.99 100 100 100 100
1% 97.61 98.83 99.16 98.91 66.74 99.42 99.35 99.06 99.98 99.94 99.23 99.3 100
2% 100 100 100 100 96.9 100 100 100 100 100 100 100 100