. 2017 Oct 29;18(6):200–205. doi: 10.1002/acm2.12210

Table 1.

WEPL deviation of simulation‐derived fitted values from measurements and adjusted R2 for simulation‐derived calibration functions and measurements‐derived calibration functions in each of the absorption media

Beam range Measurements in water‐equivalent plastic Simulated in water‐equivalent plastic Simulated in CT of water‐equivalent plastic Simulated in patient CT
WEPL deviation (mm) R2 WEPL deviation (mm) R2 WEPL deviation (mm) R2 WEPL deviation (mm)
Mean σ Max Mean σ Max Mean σ Max Mean σ Max
10 cm 0.0 0.2 0.5 0.999 0.7 0.6 1.5 0.999 0.3 0.5 0.9 0.998 0.1 0.6 1.4
15 cm 0.0 0.3 0.9 0.999 1.6 0.5 2.4 0.999 0.1 0.6 1.5 0.998 1.8 1.2 3.4
21 cm 0.0 0.4 1.1 0.999 0.9 0.6 1.3 0.999 0.8 0.6 1.6 0.999 0.5 0.9 1.7