Table 3.
Comparison of dose rate constant for different commercially available radioactive seeds
Authors | 125I seed | Material of marker | Length of marker (cm) | Dose rate constant, () | Differences between the dose rate constants (%) |
---|---|---|---|---|---|
Popescu et al | PharmaSeed BT‐125‐1 | Molybdenum rod | 0.325 | 0.950 | 0.94 |
DeMarco et al | PharmaSeed BT‐125‐2 | Silver rod | 0.325 | 0.967 | 0.83 |
TG43U1 | 6711 | Silver rod | 0.300 | 0.965 | 0.62 |
This study | GMS BT‐125‐1 | Silver rod | 0.325 | 0.959 | 0 |