Table 4.

Parameters of the IMRT test fields and gamma passing rates based on 2% DD, 2 mm DTA criteria for the 5 photon energies

Fields Description Field size (cm2) 6 MV 10 MV 18 MV 6 MV FFF 10 MV FFF
10 × 10 Absolute dose calibration 10 × 10 99.3 93.5 91.5 97.6 91.3
20 × 20 Flatness, symmetry 20 × 20 96.4 94.4 93.0 96.0 97.6
3ABUT 3 abutted segments 6 × 24 98.9 89.5 93.0 92.0 95.5
DMLC MLC offset 2 × 20 96.3 89.3 89.1 90.1 91.4
HIMRT IMRT performance Variable 99.7 94.3 94.2 99.9 94.8
HDMLC DMLC performance Variable 99.9 94.1 97.0 99.9 99.3
7SegA Picket fence 2 × 24 96.2 85.4 91.6 97.1 96.3
FourL 4 L‐shaped MLC segments Variable 93.6 76.9 85.8 91.4 90.3