Skip to main content
. 2017 Oct 29;18(6):152–168. doi: 10.1002/acm2.12205

Table 1.

Failure modes table for Gamma Knife radiosurgery

No. Step Potential failure modes Potential causes of failure Potential effects of failure O S D RPN Notes and examples of causes and failures
1 Create paper chart/eChart Incorrect patient ID data Errors in manual entry, most likely causes:
1. Omission in entry
2. Human transcription error
3. Miscommunication
Very wrong dose 2.00 8.00 1.50 24.00 Patient name is typed into the hospital database incorrectly. Information is requested from another department for a different person who actually exists. Information for the wrong patient is sent back. Suboptimal dose prescribed.
2 Previous/adjuvant treatment record Incorrect or incomplete previous treatment history 1. Personnel omission
2. Miscommunication
Very wrong dose 2.00 6.13 2.50 30.63 Historical information about patient treatment inside/outside the dept. is inaccurate or incomplete. The patient & family don't remember treatment details.
3 Physical examination Incorrect or incomplete test results 1. Lack of standardized procedures
2. Miscommunication
3. Personnel omission
Very wrong dose 2.50 3.25 1.75 14.22 Necessary physical or sensory tests are not done prior to the treatment, i.e. the most recent hearing test result is not available for an acoustic patient. Other test results may include vision, taste, motor function etc.
4 Physical examination Incorrect or incomplete patient implant info 1. Lack of standardized procedures
2. Miscommunication
Patient injury 2.13 5.38 2.13 24.27 Patient has pacemaker, dental implant etc. but the information about the implant is not available or inaccurate at the time of the treatment. MR images were acquired instead of CT. Patient injured.
5 Frame/fiducial box check Cannot start 1. Miscommunication
2. Personnel omission
Inconvenience 3.00 3.38 2.25 22.78 Patient does not show up because of a severe weather condition or a miscommunication between patient and staff. Patient shows up but finds out he/she is not on the schedule.
6 Frame/fiducial box check Use of distorted frame 1. Lack of standardized procedures
2. Personnel omission
Inaccurate volume
Inaccurate dose distribution delayed treatment
1.63 7.00 4.13 46.92 A distorted frame is used. The accuracy of the image definition/dose calculation is compromised. If the distortion is too much, patient can not be docked and the treatment can not proceed.
7 Frame/fiducial box check Bad fiducial box assembly 1. Lack of standardized procedures
2. Personnel omission
3. Inadequate training/orientation
Inaccurate volume
Inaccurate dose distribution Imaging may need to be repeated
3.38 4.88 3.75 61.70 A fiducial box is not assembled correctly, i.e. screws are loose, frontal/back or left/right pieces are switched etc.
8 Frame/fiducial box check Too much air bubble 1. Lack of standardized procedures
2. Personnel omission
Inaccurate volume
Inaccurate dose distribution
3.25 4.00 2.88 37.38 The length of the air bubbles in a MR imaging box should be minimized. Air bubbles in the fiducial box may cause certain image slices to be excluded from the image definition process.
9 Frame positioning/fixation Inappropriate local anesthesia injection Personnel inadequately trained Patient uncomfortable 3.38 4.38 3.13 46.14 Local anesthesia not given to where pins will be inserted into skull bone.
10 Frame positioning/fixation Frame not fixed firmly 1. Personnel omission
2. Personnel inadequately trained
Very wrong dose distribution Inconvenience 1.50 5.00 2.13 15.94 A frame is not attached to the patient skull firmly and becomes loose during the course of a treatment. The procedure has to start over if someone finds out about the loose frame before the treatment. Otherwise, treatment with a loose frame could mean a wrong dose distribution.
11 Frame positioning/fixation Unsterilized pin/post 1. Lack of standardized procedures
2. Personnel omission
Patient infection 1.50 3.13 1.63 7.62 The use of unsterilized pin/post may cause infection to patient and/or staff.
12 Frame positioning/fixation Inappropriate frame placement 1. Lack of training/experience
2. Miscommunication
Inconvenience (staff and patient) 2.25 6.50 1.88 27.42 The stereotactic frame is not shifted correctly when treating a peripheral lesion. A collision with the helmets (on the 4C) or the source cap (on the Perfexion) is reported from the planning system. Frame and imaging need to be repeated.
13 Frame cap/adapter test Distorted frame adapter 1. Lack of standardized procedures
2. Personnel omission
Inconvenience (staff and patient) 1.75 4.88 1.75 14.93 The frame adapter for Perfexion is bad but not found out before a treatment. Treatment can not proceed.
14 Bubble helmet measurement Inaccurate bubble measurement 1. Inadequate training of personnel
2. Human error
Inaccurate dose distribution 2.63 5.00 2.63 34.45 Bubble measurement is not accurate enough and is not noticed at the final check stage. Dose delivery is based on an inaccurate patient skull geometry.
15 Post/pin measurement Incorrect post/pin measurement 1. Inadequate training of personnel
2. Human error
Inconvenience Patient injury 1.88 5.00 1.75 16.41 Post or pin measurement is not correct. A collision is not detected by the planning system. A collision with a post or pin could stop the treatment or injure the patient on treatment.
16 Fiducial box attachment Fiducial box not attached properly 1. Inadequate training of personnel
2. Personnel omission
Inaccurate volume 2.50 3.88 2.25 21.80 The imaging fiducial box should be completely engaged with the frame and locked securely. Failure to do so may cause distortion in acquired images.
17 Patient positioning/immobilization Unsecured imaging adaptor 1. Personnel inadequately trained
2. Inadequate materials/tools
Inaccurate volume
Inaccurate dose distribution
1.63 8.63 2.75 38.54 Artifacts in images. Re‐imaging may be needed sometimes.
18 IV contrast administration Contrast not used when needed 1. Inadequate communication from MD to sim staff
2. General procedures not clearly documented
Target volume not detected 2.13 4.38 4.25 39.51 Imaging contrast is not used appropriately. May cause inaccurate definition of the treatment volume.
19 Imaging protocol selection Not enough imaging series 1. General procedures (e.g. all patients of a particular type should have certain series) not clearly documented
2. Miscommunication
Wrong target and/or OAR volume 1.75 6.38 2.38 26.50 Whole head study is not acquired, makes it difficult to check patient skull definition. Volumetric study is not acquired when needed, cause difficulty in target and/or critical organ delineation. Angio study is not acquired when needed.
20 Imaging protocol selection Wrong scan protocol used (e.g. wrong slice thickness) 1. General procedures not documented
2. Miscommunication
Inaccurate volume
inaccurate dose
2.38 5.38 2.63 33.51 Compromised CTV/OAR delineation and inaccurate dose delivery.
21 Imaging protocol selection Patient move too much during scan 1. Poor communication to the patient
2. Imaging process too long
Inaccurate dose distribution 3.00 3.25 2.25 21.94 Patient may need to be rescanned.
22 Images transferred to TPS Acquired image set accidentally deleted 1. Inattention to details
2. Inadequate training
3. Inadequate backup procedures
Inconvenience (staff and patient) 2.50 3.63 1.75 15.86 Imaging needs to be repeated. Inefficient imaging process.
23 Images transferred to TPS Incorrect image data set associated with patient 1. Pull up wrong patient's record
2. Inadequate training
Very wrong volume
Very wrong dose distribution
1.88 4.88 3.00 27.42 Patient treated on the basis of another patient's volumes.
24 Images transferred to TPS Dicom communicating failure 1. Inadequate commissioning and acceptance testing
2. Limitations of treatment planning systems or scanners
Very wrong dose distribution
Very wrong volume
1.38 7.63 2.25 23.59 A dicom communication failure causes corrupted files imported to the treatment planning system. Patient treated based on wrong information.
25 Images transferred to TPS Can not transfer image via network Defective materials (software or hardware) Inconvenience Delayed treatment 1.88 6.00 1.50 16.88 Network problem not identified and fixed before a treatment. Images can not be transferred from a scanner to a treatment planning system. CD ROM may need to be used in some cases.
26 Machine daily QA Staff not available Miscommunication Inconvenience
Patient uncomfort
Delayed treatment
3.00 1.63 2.63 12.80 After imaging, it has become clear that no or not enough coverage from the radiation oncology team is available. A patient treatment has to be cancelled after frame placement. A patient has to wait excessively long for a treatment.
27 Machine daily QA Undetected console computer malfunctioning 1. Lack of standardized procedures
2. Personnel omission
Min: Inconvenience and delayed treatment
Max: Very wrong dose
1.88 4.13 2.38 18.37 The treatment console computer is not working properly (timer, clock, system configuration etc.) but not identified during the morning QA process. Wrong clock or system configuration would prevent the system to start a treatment. Wrong timer would cause a wrong dose delivery.
28 Machine daily QA Undetected mechanical failure 1. Lack of standardized procedures.
2. Human error
Inconvenience
Delayed or partially completed treatment
2.50 5.38 3.63 48.71 There is a mechanical failure in the treatment unit (APS problem, helmet micro switch problem, helmet hoist malfunctioning, sector switch malfunctioning, couch problem etc.) but not identified during the morning QA process. Treatment can not start or may have to be stopped in the middle.
29 Create patient in TPS Wrong category (new patient/new examination) 1. Personnel omission
2. Inadequate training
Min: Confusion and inconvenience
Max: Very wrong dose
3.25 3.88 3.00 37.78 A patient underwent previous Gamma knife treatment is created as a new patient. The previous treatments could be ignored during the planning process.
30 Generate skull geometry Wrong numbers (24 point bubble measurement) Human error Inaccurate dose distribution 2.13 4.00 2.75 23.38 Bubble measurement was done correctly but wrong numbers were put into the planning system. This is not noticed during the final check. Dose calculation is based on inaccurate patient skull geometry.
31 Generate skull geometry Wrong numbers (Pin or post measurement) Human error Inconvenience Patient injury 3.25 4.38 2.13 30.21 Post or pin measurement is not put in correctly. A collision is not detected by the planning system. A collision with a post or pin could stop the treatment or injure the patient on treatment.
32 Import images Images of another patient imported Personnel omission Very wrong dose 1.63 5.88 3.25 31.03 Images of a wrong patient were imported. The images of the two patients are similar, so the error was not discovered during the planning process. Treatment is based on wrong image sets.
33 Import images Images of another patient imported Personnel omission Inconvenience 1.88 1.88 1.50 5.27 Images of a wrong patient were imported. The error was identified during the planning process. Treatment planning has to start over.
34 Import images Incomplete image sets or image series Personnel omission Inconvenience 2.38 4.63 4.13 45.31 Miss an image set, i.e., forget to import T2 volumetric study. Start to import images before the image transfer is complete. Only part of the image set is imported. This can usually be found out during the planning process.
35 Image definition/levelling Large definition error not noticed 1. Inadequate training
2. Personnel omission
Inaccurate volume Inaccurate dose distribution 3.38 4.00 3.25 43.88 The definition error for an image series is too big, either because of the use of bad equipment/protocol or excessive patient movement during the imaging process. Imaging needs to be repeated for optimal result if the large definition error is noticed.
36 Image definition/levelling Mismatch between the bubble/image not noticed 1. Inadequate training
2. Personnel omission
Inaccurate dose distribution 3.13 3.75 2.00 23.44 The skull shape from the bubble measurement and the images don't match very well. Usually bubble measurement needs to be repeated.
37 Image definition/levelling Wrong patient orientation information 1. Inadequate equipment commissioning
3. User Error
Wrong volume
Suboptimal plan
1.75 4.63 2.88 23.27 This happens most often with the definition of Angio images. The frontal/back, left/right marks were not put in correctly.
38 Image definition/levelling Inadequate leveling Inadequate training Wrong volumes 2.63 6.50 3.50 59.72 Images are not levelled appropriately. The boundary of a treatment area is blurred. Certain areas are overlooked because of poor image contrast.
39 Image fusion/registration Can not get needed image sets 1. Inadequate equipment commissioning
2. Inadequate communication
Suboptimal plan 3.25 2.63 2.13 18.13 A pre‐op/post‐op image set is needed to evaluate the progression of a previous treated (radiosurgery or surgery) disease site but there are problems transferring the images from certain image server.
40 Import previous examinations Wrong number of blue circles 1. Inadequate training
2. Personnel omission
Very wrong volumes
Very wrong dose delivery
2.63 4.75 3.25 40.52 Forget to import one of the previous examinations. Forget to turn on some of the blue circles. Cause a previously treated area with good control to be considered as a new area and treated again.
41 Image skull definition Skull not clean 1. Inadequate training
2. Personnel omission
Wrong dose distribution 2.88 3.88 2.50 27.85 The skull shape from an image skull definition is not manually edited properly. There are larger errors in certain areas of the skull contour.
42 Create treatment plan Incorrect machine configuration 1. Inadequate commissioning
2. Personnel omission
Very wrong dose
Very wrong dose distribution
1.63 3.13 3.50 17.77 Problems with the configuration of the treatment unit in the planning system. A wrong Gamma Knife unit is chosen (i.e., 4C vs Perfexion).
43 Create treatment plan Incorrect algorithm configuration 1. Inadequate commissioning
2. Miscommunication
3. Personnel omission
Very wrong dose
Very wrong dose distribution
2.75 3.00 2.00 16.50 Problem with the configuration of a dose calculation algorithm (wrong output factor, wrong CT density curve etc.).
44 Matrix definition/dose prescription Wrong location of the targets 1. Inadequate training
2. Poor communication
3. Personnel omission
Very wrong volume
Very wrong dose
2.50 3.75 2.38 22.27 Identification of the target area is not correct. Blood vessels, edemas are considered as tumors. Functional target left/right messed up, wrong location etc.
45 Matrix definition/dose prescription Overlooked targets/target area 1. Inadequate training
2. Poor communication
3. Personnel omission
Ineffective treatment 1.75 4.50 2.63 20.67 Tumors or portions of tumors are not treated because of personnel omission or poor image quality.
46 Matrix definition/dose prescription Inadequate prescription dose 1. Lack of standardized procedures.
2. Inadequate training
3. Poor communication
Very wrong dose 3.00 4.63 2.50 34.69 An inadequate prescription dose is used for a particular type of disease, depending on the experience of the planners, and/or the information gathered about the patient before the treatment.
47 Matrix definition/dose prescription Inadequate matrix size/position 1. Lack of standardized procedures.
2. Inadequate training
3. Personnel omission
Inaccurate dose calculation 3.25 3.88 2.38 29.91 A dose calculation matrix is too sparse or not centered well. Dose calculation for this target is not accurate enough.
48 Shot placement & adjustment Error in dose calc Software error Wrong dose distribution 1.38 6.50 1.63 14.52 Software cannot calculate dose correctly because of a computer problem. This can be checked by a secondary dose calculation program.
49 Contouring Wrong organ, wrong site 1. Inadequate training
2. Personnel omission
Very wrong volumes 2.75 3.88 1.88 19.98 Wrong target volume or critical organ contours lead directly to very wrong dose distributions and volumes.
50 Contouring Poorly drawn contours (spikes, sloppy, etc.) Inattention Suboptimal plan (worst case wrong dose distribution) 2.38 4.38 2.13 22.08 Drawings generally correct but have inappropriate spikes, sharp corners, etc. Plan evaluation is based on the imperfect contours. Wrong dose distribution in worst case.
51 Contouring/plan evaluation Needed contour not drawn 1. Inadequate training
2. Inattention, lack of time
3. Failure to review own work
Suboptimal plan (worst case wrong dose distribution)
Inconvenience
4.88 2.63 2.75 35.19 A critical structure is not drawn when needed. The structure is not included in plan evaluation. The plan is not optimal. Make it difficult to make clinical decisions for future treatments. Examples include skin dose is not evaluated, optical structure doses are not documented.
52 Plan evaluation Inadequate evaluation 1. Not enough time/effort spent
2. Inadequate training
3. Poor evaluation strategy
Wrong dose
Wrong dose distribution
1.63 6.88 2.38 26.53 One needs to look at DVH and dose distribution on a slice by slice basis. But because of poor training or not spending enough time to evaluate the DVH or isodose distribution, the result can be an inadequate evaluation of the plan.
53 Plan approval Wrong patient Inattention Very wrong dose
Very wrong dose distribution
1.75 6.50 4.13 46.92 Exported a treatment plan for a wrong patient. This happens rarely and can be found out in most of the cases.
54 Plan approval Wrong plan approved 1. Miscommunication
2. Inattention
3. Inadequate procedure
Very wrong dose
Very wrong dose distribution
2.25 5.88 3.63 47.92 Exported one of the trial plans rather than the final plan.
55 Plan approval Matrix size not fine‐tuned 1. Inadequate procedure
2. Procedure not followed
3. Personnel omission
Inaccurate dose distribution 1.75 4.63 1.88 15.18 The sizes of the matrices are not adjusted appropriately before the plan is approved, resulting in inaccurate dose calculations.
56 Plan approval Collision check not performed 1. Inattention
2. Lack of procedure
3. Procedure not followed
Inconvenience
Patient injury
1.88 4.88 1.75 16.00 There is a collision between the helmet/cap and the patient skull, frame post/pin. Treatment may be interrupted and replan maybe needed. A collision with a pin or skull could injury the patient as well.
57 Plan approval Setup not checked 1. Inattention
2. Lack of procedure
3. Procedure not followed
Inconvenience
Suboptimal plan
2.63 5.00 2.50 32.81 Setup/shot summary is not checked. The exported plan contains an unwanted Gamma angle or helmet change, a shot with very short treatment time etc.
58 Chart finish‐up Not signed appropriately 1. Inattention
2. Miscommunication
Inconvenience 3.25 2.25 1.75 12.80 The hard copy of the treatment plan is not signed appropriately. It could also mean that the final plan has not been checked and agreed by all planners in some cases.
59 Chart finish‐up Unnoticed plugged pattern 1. Miscommunication
2. Lack of standard procedures
Very wrong dose
Very wrong dose distribution
Very wrong volume
1.88 5.13 2.13 20.42 A plug pattern is used on 4C but the hard copy does not show it, either because of an incorrect printout selection or a breakdown of the printer. Treatment proceeded without the plug pattern.
60 Secondary dosimetry check Undetected planning computer failure 1. Lack of procedure
2. Inadequate procedure
3. Procedure not followed
Very wrong dose
Very wrong dose distribution
Very wrong volume
1.50 7.38 3.13 34.57 A failure in the treatment planning system gives an erroneous delivery time but this not caught by the second check. Or the secondary check is simply not done.
61 Plan export for treatment Wrong plan exported Personnel omission Very wrong dose;
Very wrong dose distribution
Very wrong volume
1.13 9.13 1.75 17.96 This is highly unlikely for the recent versions of the planning system because only an approved plan can be exported. There are not many reasons to have several outstanding approved plans at a time in the Gamma Knife planning system.
62 Load plan in treatment console Communication failure 1. Cable pulled out accidentally
2. Bad wire
3. Network problem
Inconvenience 2.13 1.88 2.63 10.46 The exported plan can not get to the treatment console computer. Treatment can not start. The communication between the planning computers and the treatment console computer is usually not checked during the morning QA.
63 Load plan in treatment console Wrong plan loaded Personnel omission Very wrong everything 1.13 7.00 3.50 27.56 Load a plan for a different patient. This is also highly unlikely because there are not many reasons to keep more than one plan in the treatment queue.
64 Patient ID check Incorrect patient in the room 1. Lack of standard procedures
2. Poorly trained personnel
Very wrong everything 1.88 3.75 2.50 17.58 Inadequate check.
65 Frame adapter attachment Frame adapter not attached properly 1. Inadequate training/orientation
2. Personnel omission
Inconvenience
Inaccurate dose delivery
2.75 7.75 5.75 122.55 On the Perfexion, the frame adaptor should be attached properly. Failure to do so may result in an interruption in the treatment process or an imperfect dose delivery. Elekta has a field notice about this.
66 Patient positioning Couch vertical position not ideal 1. Lack of standard procedures
2. Inattention to detail.
3. Poorly trained personnel
Patient uncomfortable 2.25 3.75 3.50 29.53 The couch vertical position should be optimized so that the patient neck does not get stretched much. This is important for long treatments. The couch vertical position may need to be adjusted before each run. A different run means a different Gamma angle or tumors in a different area are being treated.
67 Lock patient Incorrect patient setup Personnel omission Very wrong dose distribution 1.88 4.13 2.00 15.47 Patient setup with wrong stereotactic coordinates in the trunnion mode. Patient setup at a wrong Gamma angle. On the Perfexion treatment can not start with a wrong Gamma angle.
68 Lock patient Patient head position not firmly fixed 1. Bad equipment
2. Personnel omission
Wrong dose distribution 2.63 3.63 2.50 23.79 It is good practice to check if the patient head is fixed properly before a treatment. Equipment failure, untighten or loose screws may cause unsecured patient head position. This is more important for the trunnion mode on the 4C.
69 Clearance test Undetected collision 1. Personnel omission
2. Procedure not followed
Interrupted treatment
Suboptimal plan
Inconvenience Patient injury
1.50 6.50 1.75 17.06 A necessary clearance test is not performed or not performed properly. A treatment is interrupted because of a collision. Replan is needed. Patient may get injured.
70 Final check Stuff on the couch 1. Personnel omission
2. Lack of standard procedure
3. Procedure not followed
Interrupted treatment 2.50 5.38 2.88 38.63 Pillows, blankets, screw drivers etc. are left on the couch. Treatment interrupted.
71 Final check Tubes/cables not long enough 1. Personnel omission
2. Lack of standard procedure
Interrupted treatment
Patient uncomfortable
2.88 2.75 3.50 27.67 A tube/wire (oxygen, pulse monitor, anesthesia equipment etc.) is attached to a patient. The tube/wire gets pulled when the couch moves into the treatment position. Treatment interrupted.
72 Final check Wrong plug pattern 1. Personnel omission
2. Lack of double check
Wrong dose distribution 1.75 9.00 2.88 45.28 The plug pattern is not put in correctly on 4C.
73 Final check Inserts/Plugs not secured 1. Personnel omission
2. Lack of double check
Interrupted treatment 3.25 5.13 3.63 60.38 An insert or plug is not firmly attached to a helmet. The insert/plug drops during a treatment. Cause the treatment to stop.
74 Final check Bolus not used when needed. 1. Inadequate training
2. Inadequate procedure
3. Procedure not followed
Wrong or very wrong dose distribution 3.00 3.88 2.50 29.06 Bolus should be used for the treatment of superficial lesions. The surface portion of the lesion may be underdose if a bolus is not applied.
75 Beam on Can not start treatment 1. Lack of training
2. Personnel omission
Delayed treatment 2.25 2.88 2.38 15.36 All the interlocks (couch release handle, side pieces, emergency buttons) has to be cleared before the beam‐on button lightens up. A patient is on the table but there is an un‐cleared interlock.
76 Beam on Patient arms collide with machine 1. Lack of communication with patient
2. Inattention
Interrupted treatment
Patient uncomfortable or injury
1.50 6.25 3.38 31.64 This happens most often when a restless or confused patient is being moved in or out of the treatment unit.
77 Treatment log book Forget to record start/end time Inattention No effect on patient Inconvenience 2.50 2.25 2.25 12.66 Incomplete documentation.
78 Treatment monitoring Treated shots not recorded correctly 1. Lack of standard procedure
2. Procedure not followed
3. Personnel omission
Very wrong dose
Very wrong dose distribution
2.00 5.50 2.50 27.50 Delivered shots are not followed closely and recorded correctly on the treatment copy. Cause confusion in the shots that have been delivered. This happens more likely on 4C with trunnion shots.
79 Treatment monitoring Patient hand reaches up 1. Lack of communication with patient
2. Inattention
Inaccurate dose delivery Interrupted treatment 3.00 2.63 3.38 26.58 Patient hands in the beam pathways. Cause underdose from some beam angles. Patient hand touch the source cap. Cause the treatment to be interrupted.
80 Treatment monitoring Inappropriate sedation Lack of experience/training Patient comfort 1.88 6.75 1.88 23.73 A patient is overly sedated. Cause the treatment to be interrupted.
81 Treatment monitoring Anesthesia problem Lack of experience/training Interrupted treatment 2.75 2.75 1.75 13.23 A patient is treated under general anesthesia. But an anesthesia problem cause a treatment to be interrupted.
82 Treatment monitoring Poor choice of treatment sequence Poor user choice Ineffective treatment process 2.00 4.88 2.00 19.50 On 4C, when multiple helmets, multiple Gamma angles, high/low docking are involved, it is necessary to figure out a treatment sequence that minimizes the treatment time.
83 Treatment monitoring Manmade treatment interruption Inattention Interrupted treatment 2.00 1.63 3.13 10.16 During the course of a treatment, someone accidently press the power button, door open, emergency, pause button, or spill liquids on the treatment console computer, cause the treatment to be interrupted.
84 Treatment monitoring Unfinished treatment Personnel omission Very wrong dose 2.38 4.63 2.63 28.83 A treatment run is completely ignored.
85 Follow up phone calls Unaddressed patient concerns Inadequate procedure
Personnel omission
Patient comfort 1.88 1.75 1.75 5.74 Patient has question to ask after Gamma Knife treatment, but is not answered appropriately.
86 Follow up scan Follow up scan not at the right time Miscommunication to patient Patient care 2.50 1.88 1.63 7.62 May cause delay in further treatment.