Skip to main content
. 2017 Oct 29;18(6):152–168. doi: 10.1002/acm2.12205

Table 2.

Failure modes that are treatment machine dependent

No. Step Potential failure modes Potential causes of failure Potential effects of failure Machine Notes and examples of causes and failures
13 Frame cap/adapter test Distorted frame adapter 1. Lack of standardized procedures
2. Personnel omission
Inconvenience (staff and patient) Perfexion The frame adapter for Perfexion is bad but not found out before a treatment. Treatment can not proceed.
65 Frame adapter attachment Frame adapter not attached properly 1. Inadequate training/orientation
2. Personnel omission
Inconvenience
Inaccurate dose delivery
Perfexion On the Perfexion, the frame adaptor should be attached properly. Failure to do so may result in an interruption in the treatment process or an imperfect dose delivery. Elekta has a field notice about this.
59 Chart finish‐up Unnoticed plugged pattern 1. Miscommunication
2. Lack of standard procedures.
Very wrong dose
Very wrong dose distribution
Very wrong volume
4C A plug pattern is used on 4C but the hard copy does not show it, either because of an incorrect printout selection or a breakdown of the printer. Treatment proceeded without the plug pattern.
72 Final check Wrong plug pattern 1. Personnel omission
2. Lack of double check
Wrong dose distribution 4C The plug pattern is not put in correctly on 4C.
73 Final check Inserts/Plugs not secured 1. Personnel omission
2. Lack of double check
Interrupted treatment 4C An insert or plug is not firmly attached to a helmet. The insert/plug drops during a treatment. Cause the treatment to stop.
82 Treatment monitoring Poor choice of treatment sequence Poor user choice Ineffective treatment process 4C On 4C, when multiple helmets, multiple Gamma angles, high/low docking are involved, it is necessary to figure out a treatment sequence that minimizes the treatment time.
12 Frame positioning/fixation Inappropriate frame placement 1. Lack of training/experience
2. Miscommunication
Inconvenience (staff and patient) Both The stereotactic frame is not shifted correctly when treating a peripheral lesion. A collision with the helmets (on the 4C) or the source cap (on the Perfexion) is reported from the planning system. Frame and imaging need to be repeated.
67 Lock patient Incorrect patient setup Personnel omission Very wrong dose distribution Both Patient setup with wrong stereotactic coordinates in the trunnion mode. Patient setup at a wrong Gamma angle. On the Perfexion treatment can not start with a wrong Gamma angle.
68 Lock patient Patient head position not firmly fixed 1. Bad equipment
2. Personnel omission
Wrong dose distribution Both It is good practice to check if the patient head is fixed properly before a treatment. Equipment failure, untighten or loose screws may cause unsecured patient head position. This is more important for the trunnion mode on the 4C.
78 Treatment monitoring Treated shots not recorded correctly 1. Lack of standard procedure
2. Procedure not followed
3. Personnel omission
Very wrong dose
Very wrong dose distribution
Both Delivered shots are not followed closely and recorded correctly on the treatment copy. Cause confusion in the shots that have been delivered. This happens more likely on 4C with trunnion shots.