Skip to main content
. 2017 Oct 29;18(6):152–168. doi: 10.1002/acm2.12205

Table 4.

Failure modes with the highest severity scores for Gamma Knife Radiosurgery

No. Step Potential failure modes Potential causes of failure Potential effects of failure S Notes and examples of causes and failures
61 Plan export for treatment Wrong plan exported Personnel omission Very wrong dose
Very wrong dose distribution
Very wrong volume
9.13 This is highly unlikely for the recent versions of the planning system because only an approved plan can be exported. There are not many reasons to have several outstanding approved plans at a time in the Gamma Knife planning system.
72 Final check Wrong plug pattern 1. Personnel omission
2. Lack of double check
Wrong dose distribution 9.00 The plug pattern is not put in correctly.
17 Patient positioning/immobilization Unsecured imaging adaptor 1. Personnel inadequately trained
2. Inadequate materials/tools
Inaccurate volume
Inaccurate dose distribution
8.63 Artifacts in images. Re‐imaging may be needed sometimes.
1 Create paper chart/eChart Incorrect patient ID data Errors in manual entry, most likely causes:
1. Omission in entry
2. Human transcription error
3. Miscommunication
Very wrong dose 8.00 Patient name is typed into the hospital database incorrectly. Information is requested from another department for a different person who actually exists. Information for the wrong patient is sent back. Suboptimal dose prescribed.
65 Frame adapter attachment Frame adapter not attached properly 1. Inadequate training/orientation
2. Personnel omission
Inconvenience
Inaccurate dose delivery
7.75 On the Perfexion, the frame adaptor should be attached properly. Failure to do so may result in an interruption in the treatment process or an imperfect dose delivery. Elekta has a field notice about this.