Skip to main content
. 2017 Feb 2;18(2):44–49. doi: 10.1002/acm2.12043

Table 1.

Comparison of dosimetric parameters of lateral dose profiles at mid‐range of SOBPs in a Solidwater® phantom

SOBP Penumbra 20%–80% (mm) Half‐width of 95% shoulder (mm) Half‐width of 5% shoulder (mm)
Meas AcurosPT Topas Meas AcurosPT Topas Meas AcurosPT Topas
R120M40 8.0 8.3 8.3 41.7 40.8 40.8 58.3 59.2 59.2
R200M100 7.8 8.1 8.0 39.7 38.5 39.2 56.5 57.9 57.3
R305M100 10.2 11.8 10.7 37.4 34.6 36.0 59.8 63.3 61.0