Thoracic ion chamber measurements in vertebral body

Treatment Plan Ion Chamber Measurement (Gy) Treatment Platform
TomoTherapy Vero TrueBeam (Flattened) TrueBeam (FFF) CyberKnife
Plan A Calculated Dose 16.5 20.6 16.5 16.3 17.9
Measured Dose 16.4 21.1 17.0 16.5 18.5
% Difference ‐0.3 2.6 3.1 1.4 3.0
Plan B Calculated Dose 16.6 20.8 16.4 16.5 21.8
Measured Dose 16.6 21.5 16.3 16.6 22.4
% Difference 0.0 3.2 ‐0.3 0.6 2.4
Plan D Calculated Dose 16.2 21.2 16.0 16.2 20.4
Measured Dose 16.5 21.5 16.1 16.3 20.6
% Difference 2.2 1.4 0.6 0.8 0.7