. 2015 Jan 8;16(1):170–182. doi: 10.1120/jacmp.v16i1.5120

Table 3.

Lumbar measurements with GAFCHROMIC EBT3 film with gamma analysis utilizing the dose criterion relative to the pixel under analysis method

Treatment Plan Local γ‐analysis constraints w/ 10% threshold Treatment Platform
TomoTherapy Vero TrueBeam (Flattened) TrueBeam (FFF) CyberKnife
Plan A 3%/3 mm 100.0 99.9 100.0 100.0 99.5
2%/2 mm 98.7 98.3 96.3 99.5 98.9
1%/1 mm 71.7 49.8 64.7 72.8 84.2
Plan B 3%/3 mm 100.0 99.9 100.0 100.0 96.1
2%/2 mm 99.9 96.0 97.8 98.9 93.3
1%/1 mm 85.5 51.3 71.7 79.7 74.3
Plan C 3%/3 mm 100.0 100.0 100.0 100.0 99.2
2%/2 mm 99.5 100.0 100.0 100.0 97.5
1%/1 mm 77.1 47.2 81.3 79.2 53.2
Plan D 3%/3 mm 99.5 100.0 100.0 100.0 97.5
2%/2 mm 96.7 99.8 96.2 99.6 95.1
1%1 mm 80.7 75.5 81.8 82.1 79.6