Table 5.

Comparison of EBT3 vs. MD‐V3 GAFCHROMIC film for gamma analysis of TomoTherapy plan B using both dose criteria relative to maximum and also to the pixel in question

Treatment Plan γ‐analysis constraints w/ 10% threshold GAFCHROMIC Film
EBT3 MD‐V3
TomoTherapy Plan B 3%/3 mm 100 / 100 100 / 100
2%/2 mm 100 / 99.9 100 / 99.9
1%/1 mm 97.4 / 91.9 98.9 / 84.5