. 2015 Jul 8;16(4):125–148. doi: 10.1120/jacmp.v16i4.5313

Table 8.

Composite Gafchromic film and ion chamber results for the measurements in the high‐dose and low‐dose region for both IMRT and RapidArc plans

6XFFF 10X FFF
Plan Global Gamma
3%/3 mm
Point Dose (Percent Difference) Global Gamma
3%/3 mm
Point Dose (Percent Difference)
Hard C IMRT (PTV) 90.6 3.5% 90.0 4.7%
Hard C IMRT (low dose) 87.4 2.4% 91.1 11.9%
Hard C RA (PTV) 93.0
0.5%
97.5
0.1%
Hard C RA (low dose) 95.2
4.2%
98.7 2.4%
HN IMRT (PTV) 94.1 1.0% 98.1 2.5%
HN IMRT (low dose) 97.0 0.4% 99.5 1.3%
HN RA (PTV) 97.9 0.4% 98.7
0.4%
HN RA (low dose) 98.1 0.2% 98.4
0.2%
HN SIB IMRT (PTV) 97.5
0.9%
98.9 1.0%
HN SIB IMRT (low dose) 98.6
0.2%
97.5 1.7%
HN SIB RA (PTV) 99.0
0.4%
98.1
0.7%
HN SIB RA (low dose) 98.1 0.7% 97.8 0.5%
Prostate IMRT (PTV) 95.7
1.8%
98.2
0.1%
Prostate IMRT (low dose) 89.6
3.1%
95.3
0.2%
Prostate RA (PTV) 99.1
1.0%
99.3 0.0%
Prostate RA (low dose) 95.7
3.1%
99.4
1.9%
Prostate LN IMRT (PTV) 86.2
0.9%
99.0
0.5%
Prostate LN IMRT (low dose) 96.2
0.8%
98.9
0.2%
Prostate LN RA (PTV) 98.7 0.6% 99.2 0.4%
Prostate LN RA (low dose) 99.0 1.7% 97.9 1.6%
SIMT RA (low dose) 98.9 2.9% 100.0
0.7%