Table 9.
Summary of IROC phantom irradiation results for the lung and spine phantoms. The dosimetric precision of TLD is 3%
Phantom | TLD Location | IROC vs. Inst | Criteria | Film Plane | Gamma Index | Criteria | |
---|---|---|---|---|---|---|---|
Lung Phantom | PTV TLD sup | 0.97 | 0.92–1.08 | Axial | 100% |
|
|
PTV TLD inf | 0.98 | 0.92–1.08 | Coronal | 100% |
|
||
Sagittal | 100% |
|
|||||
Spine Phantom | PTV TLD sup ant | 1.01 | 0.93–1.07 | Axial | 90% |
|
|
PTV TLD inf ant | 1.00 | 0.93–1.07 | |||||
PTV TLD sup post | 1.00 | 0.93–1.07 | Sagittal | 91% |
|
||
PTV TLD inf post | 0.99 | 0.93–1.07 |