. 2016 Jul 8;17(4):391–401. doi: 10.1120/jacmp.v17i4.6231

Table 1.

The characteristics of our study

Facility 1 2 3 4 5 6 7 8 9 10 11 12 13
X‐ray System A Infinie Celeve
i8000V
TOSHIBA
B Infinti Celeve
i8000V
TOSHIBA
C Bransist Safire SHIMADZU D Advantx GE E Infinti Celeve DP TOSHIBA F Axiom zee biplane SIEMENS G Infinti Celeve
i8000V
TOSHIBA
H Infinti Celeve
i8000V
TOSHIBA
I Infinie Celeve
i8000C
TOSHIBA
J Infinie Celeve
i8000F
TOSHIBA
K Axiom zee biplane SIEMENS L Axiom Artis dBC SIEMENS M Trinias B‐12 SHIMADZU N Axiom Artis BA SIEMENS O Bransist Safire SHIMADZU P Infinix Celeve
i8000V
TOSHIBA
Q Allura Xper FD20 PHILIPS R Allura Xper FD10 PHILIPS
Image receptor FPD FPD FPD I.I. FPD FPD FPD FPD FPD FPD FPD FPD FPD I.I. FPD FPD FPD FPD
Actual diameter [cm] 16.4 16.4 18.0 14.5 16.4 13.5 11.6 11.6 14.0 18.5 13.5 13.5 14.3 14.2 14.5 18.6 14.9 14.9
Years of use 2.6 2.6 4.2 12.0 8.7 1.2 7.6 2.0 0.5 6.8 1.5 7.8 0.5 10.0 5.5 7.0 5.0 10.0
Total filtration [mm] 2.5 Al 2.5 Al 1.5 Al 2.5 Al 2.8 Al 2.5 Al 2.5 Al 2.5 Al 2.5 Al 2.5 Al 2.5 Al 2.5 Al 1.5 Al 2.5 Al 1.5 Al 2.3 Al 2.5 Al 2.5 Al
Grid ratio [:1] 13 13 10 10 13 15 13 13 13 13 15 15 13 17 10 13 15 13
Fluoroscopy
Tube potential [kV] 80.0 80.0 77.0 80.0 80.0 90.8 70.0 70.0 70.0 70.0 68.4 70.0 76.0 77.0 73.0 80.0 91.0 78.0
Pulse rate [pulses/s] 10.0 10.0 12.5 Cont. 15.0 7.5 7.5 7.5 7.5 7.5 7.5 7.5 15.0 7.5 15.0 15.0 15.0 15.0
Additional filtration [mm] 0.3 Cu 0.3 Cu
1.5Al+0.6Cu
0.06 Ta 0.9 Cu 0.2 Cu 0.2 Cu 0.2 Cu 0.2 Cu 0.2 Cu 0.2 Cu
1.5Al+0.6Cu
0.9 Cu
1.5Al+0.3Cu
0.06 Ta
1.0Al+0.4Cu
0.4 Cu
Entrance dose [mGy/min] 10.5 10.8 8.8 22.1 21.9 2.1 18.8 18.5 18.7 8.8 8.3 9.3 12.5 5.2 28.2 12.2 17.1 16.8
Pulse dose [mGy/pulse] 0.018 0.018 0.012 0.012 0.024 0.005 0.042 0.041 0.042 0.020 0.018 0.021 0.014 0.012 0.031 0.014 0.019 0.019
Image quality score
Spatial resolution 3.0 3.5 3.0 3.5 4.0 1.0 5.0 6.0 6.0 5.0 3.5 4.0 4.0 3.0 4.0 4.0 4.0 4.0
Low contrast detectability 2.0 2.0 1.5 1.5 2.0 0.0 4.0 4.0 4.5 2.5 2.0 3.0 2.0 2.5 4.0 2.0 2.0 2.0
Dynamic range 3.5 4.0 3.5 3.0 5.0 1.5 5.0 5.5 5.0 5.0 4.0 4.0 5.5 4.0 5.0 5.0 5.0 3.5
Cineradiography
Tube potential [kV] 74.0 73.0 68.0 80.0 74.0 76.0 74.0 73.0 74.0 74.0 72.4 68.3 70.0 76.2 68.0 71.0 68.0 71.0
Frame rate [frames/s] 15.0 15.0 12.5 12.5 15.0 15.0 15.0 15.0 15.0 15.0 10.0 10.0 15.0 15.0 15.0 15.0 15.0 15.0
Additional filtration [mm] 1.8 Al 1.8 Al 1.0 Al 0.03 Ta 0.3 Cu 1.8 Al 1.8 Al 1.8 Al 1.8 Al 0.1 Cu 0.1 Cu
1.0Al+0.01Au
0.1 Cu
1.5Al+0.01Au
1.8 Al
1.0Al+0.4Cu
0.1 Cu
Entrance dose [mGy/lOs] 16.3 16.6 24.7 28.1 44.5 6.3 41.3 37.8 49.3 17.6 5.1 9.9 19.1 23.2 27.2 29.5 27.1 19.4
Frame dose [mGy/frame] 0.109 0.111 0.198 0.225 0.297 0.042 0.275 0.252 0.329 0.117 0.051 0.099 0.127 0.155 0.181 0.197 0.181 0.129
Image quality score
Spatial resolution 5.0 5.0 5.0 5.0 5.0 4.0 6.0 7.0 6.0 6.0 5.0 5.0 5.0 4.5 5.0 5.0 5.0 5.0
Low contrast detectabiliry 3.0 3.5 4.0 3.5 4.0 2.0 5.0 5.5 4.5 4.0 3.5 4.0 3.0 4.0 4.5 3.5 3.5 4.0
Dynamic range 5.0 6.5 7.0 3.5 7.0 4.0 10.5 10.0 7.0 7.5 4.5 4.5 7.0 6.0 7.5 6.5 6.5 6.5

FPD=flatpaneldetector; I.I.=image intensifier; Cu=copper; Al=aluminium; Ta=tantalum; Au=gold.