Skip to main content
. 2016 Jul 8;17(4):48–58. doi: 10.1120/jacmp.v17i4.5965

Table 2.

MC‐to‐PB ratios of PTV V100 for virtual phantom cases with spherical GTVs of 15 mm, 30 mm, and 50 mm diameters, planned individually with 90%, 80%, 70%, and 60% prescription isodose lines and with 6 MV photon beams, as based on PB calculation

V100(MC)/V100(PB)
RxLine=90%
RxLine=80%
RxLine=70%
RxLine=60%
DGTV=15mm
0.01 0.09 0.17 0.27
DGTV=30mm
0.23 0.31 0.47 0.57
DGTV=50mm
0.48 0.54 0.70 0.78