Table 2.

Summary data of CBCT scans performed on lung cancer patients, including the initial angle of the treatment gantry, the duration of CBCT/AlignRT synchronization, the number of breathing cycles analyzed per patient, and the corresponding cycle length

Patient Angular Accuracy (°) Real CBCT Starting Angle (°) Duration of Synchronization (s) Number of Breathing Cycles Cycle Length (s)
P1 0.03
178.07
30.3 9 2.8
P2
0.10
178.53
31.9 5 4.6
P3 0.09
178.50
19.2 9 1.8
P4 0.05
178.60
34.5 8 3.3
P5 0.20
178.27
26.3 5 3.5
P6 0.04
178.38
29.0 6 3.3
P7 0.06
178.61
24.6 5 3.2