. 2015 Mar 8;16(2):87–98. doi: 10.1120/jacmp.v16i2.5088

Table 3.

Warning and Action thresholds for planar radiographic modes based on run chart analysis

kV MV
Warning Action Warning Action
f30
a
2% 4%
f30
2% 4%
f40
a
1% 3%
f40
1% 3%
f50
a
1% 3%
f50
1% 3%
Dose (μGy) 1% 2% Dose (μGy) 1% 2%
Energy (kV) 1% 2%
a

fx parameters measured in units of (lp/mm); sample size of 60 measurements taken over four‐month time period.