Warning and Action thresholds for full‐fan kV CBCT mode based on run chart analysis

Warning Action
Uniformity (HU/HU) 6% 12%
Noise (HU) 6% 12%
Spatial Resolution (lp/mm)
f30
9% 18%
f40
9% 18%
f50
8% 16%
HU Constancy (HU)
Lung 3% 6%
Water 6% 12%
Bone 2% 4%
Geometric Distortion (mm)
AP 1% 2%
LAT 1% 2%
Slice thickness (mm) 6% 12%
Dose (mGy)
Center 1% 2%
Periphery 1% 2%