Skip to main content
. 2015 Mar 8;16(2):87–98. doi: 10.1120/jacmp.v16i2.5088

Table 6.

Institutional imaging QA threshold tolerances for MV/kV planar radiographic modes

Frequency Quality Metric Quality Check Tolerance Level (Percent of Baseline)
Warning Action
Monthly High contrast spatial resolution (1p/mm)
f30
>2%
>4%
f40
>2%
>4%
f50
>2%
>4%
Annual Imaging quality Dose Energy
>1%
>2%
Energy
>1%
>2%