Skip to main content
. 2015 May 8;16(3):329–338. doi: 10.1120/jacmp.v16i3.5427

Table 1.

Gamma values calculated by Dosimetry Check system with fluence from EPID, seven29, and MatriXX on 15 IMRT plans.

Anatomical Site EPID seven29 Matrixx
Abdomen 98.85% 98.08% 94.90%
Brain 1 97.77% 95.09% 93.51%
Brain 2 94.03% 94.30% 95.03%
Brain 3 99.26% 98.45% 97.70%
Head and Neck 1 92.51% 92.68% 90.62%
Head and Neck 2 98.59% 93.74% 96.33%
Head and Neck 3 96.86% 96.48% 95.56%
Liver 1 99.09% 97.93% 98.22%
Liver 2 99.62% 98.90% 99.13%
Liver 3 99.46% 97.61% 97.13%
Lung 1 93.71% 92.19% 93.64%
Lung 2 95.66% 92.34% 93.38%
Lung 3 96.12% 95.59% 93.59%
Prostate 1 93.85% 92.60% 92.61%
Prostate 2 96.28% 94.82% 93.68%
Average 96.78% 95.39% 95.00%
SD 2.41% 2.40% 2.34%