Skip to main content
. 2015 May 8;16(3):60–79. doi: 10.1120/jacmp.v16i3.4636

Table 1.

Analysis of systematic errors and statistical dispersions in water, for a field of 6×6cm2 at 18 MeV and SSD=100cm. The statistical accuracy was 2% and the grid size was 2×2mm2.

Curve Region Maximum Systematic Error Maximum Statistical Dispersion
dmax – d80%
0.3%
1.5%
d80% – d20% −0.4 mm −0.8 mm
d20% – end 0.2% 1.0%